Recall Z-0562-2016— Beckman Coulter Inc.

Class II · Terminated

Class II FDA medical device recall by Beckman Coulter Inc., initiated 2015-12-01, terminated 2016-07-29. FDA's recall record names no specific 510(k) clearance for this event. Product: MicroScan Pos Breakpoint Combo Panel Type 23, Part No. B1017-206 MicroScan Pos (Positive) panels are designed for use in determining antimicrobial agent susceptibility and/or identification to the species level of rapidly growing aerobic and facultative gram-positive cocci, some fastidious aerobic gram-positive cocci and Listeria monocytogenes.. Reason: Beckman Coulter has confirmed falsely negative results for Voges-Proskauer (VP) (pale-pink/brown/colorless) with Quality Control (QC) American Type Culture Collection (ATCC) organism Staphylococcus aureus ATCC 29213 in a portion of affected lots of MicroScan Pos Combo and MicroScan Pos Breakpoint Combo panels. The expected result is positive..

Recalling firm
Beckman Coulter Inc.
Classification
Class II
Status
Terminated
Initiated
2015-12-01
Terminated
2016-07-29
Firm location
Brea, CA, United States

Product description

MicroScan Pos Breakpoint Combo Panel Type 23, Part No. B1017-206 MicroScan Pos (Positive) panels are designed for use in determining antimicrobial agent susceptibility and/or identification to the species level of rapidly growing aerobic and facultative gram-positive cocci, some fastidious aerobic gram-positive cocci and Listeria monocytogenes.

Reason for recall

Beckman Coulter has confirmed falsely negative results for Voges-Proskauer (VP) (pale-pink/brown/colorless) with Quality Control (QC) American Type Culture Collection (ATCC) organism Staphylococcus aureus ATCC 29213 in a portion of affected lots of MicroScan Pos Combo and MicroScan Pos Breakpoint Combo panels. The expected result is positive.

Data sourced from openFDA. This site is unofficial and independent of the FDA.