Recall Z-0557-2022— Covidien Llc

Class II · Ongoing

Class II FDA medical device recall by Covidien Llc, initiated 2021-12-17. It names one FDA 510(k) clearance: K163231. Product: Chameleon PTA Balloon Dilation Catheter 8 mm x 40 mm, Chameleon PTA Balloon Catheter with Injection Port 8 mm x 40 mm; Cardiovascular. Product Numbers CH08-40-75GB and CH08-40-75US. Reason: The firm has received six (6) complaint reports from customers indicating difficulty removing the balloon protector sleeve from the catheter (e.g., excessive force required to remove the protector sleeve). If the balloon is damaged while removing the protector sleeve and the catheter is subsequently used in a patient procedure, patient harm may potentially occur, including perforation of vessels, foreign body in patient, embolism, risks associated with surgical intervention, or delay to treatment/therapy..

Recalling firm
Covidien Llc
Classification
Class II
Status
Ongoing
Initiated
2021-12-17
Firm location
Mansfield, MA, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

Chameleon PTA Balloon Dilation Catheter 8 mm x 40 mm, Chameleon PTA Balloon Catheter with Injection Port 8 mm x 40 mm; Cardiovascular. Product Numbers CH08-40-75GB and CH08-40-75US

Reason for recall

The firm has received six (6) complaint reports from customers indicating difficulty removing the balloon protector sleeve from the catheter (e.g., excessive force required to remove the protector sleeve). If the balloon is damaged while removing the protector sleeve and the catheter is subsequently used in a patient procedure, patient harm may potentially occur, including perforation of vessels, foreign body in patient, embolism, risks associated with surgical intervention, or delay to treatment/therapy.