Recall Z-0554-2025— Shimadzu Corp. Analytical & Measuring Instruments Div.

Class II · Ongoing

Class II FDA medical device recall by Shimadzu Corp. Analytical & Measuring Instruments Div., initiated 2024-10-09. FDA's recall record names no specific 510(k) clearance for this event. Product: Autosampler, Models: SIL-40C XR CL, SIL-40C X3 CL. Reason: There have been cases of failure of the main board due to condensation water occurring in the gap between thermal insulation material on the underside of the main base dropping onto the main board The symptoms and errors differ depending on the position where the water drops adhere. To date there have been examples of reports of inter-board communication errors, HOME POSITION error or SLIP error of XYZ drive units. The failure could result in delay of patient sample testing..

Recalling firm
Shimadzu Corp. Analytical & Measuring Instruments Div.
Classification
Class II
Status
Ongoing
Initiated
2024-10-09
Firm location
Kyoto, Japan

Product description

Autosampler, Models: SIL-40C XR CL, SIL-40C X3 CL

Reason for recall

There have been cases of failure of the main board due to condensation water occurring in the gap between thermal insulation material on the underside of the main base dropping onto the main board The symptoms and errors differ depending on the position where the water drops adhere. To date there have been examples of reports of inter-board communication errors, HOME POSITION error or SLIP error of XYZ drive units. The failure could result in delay of patient sample testing.