Recall Z-0553-2016— GE Medical Systems, LLC

Class II · Terminated

Class II FDA medical device recall by GE Medical Systems, LLC, initiated 2015-12-07, terminated 2018-11-08. It names one FDA 510(k) clearance: K110610. Product: GE Healthcare, 3.0T 6 Channel Phased Array Flex Coil, M0050SS. Product Usage The 3.0T GE 6-Channel Phased Array Flex Coil is indicated for use on the order of the physician in conjunction with a 3.0T MRI scanner, as an accessory to produce 2D and 3D images.. Reason: Coil overheating can occur when the device is used in Mode 2 setup. This could lead to a serious patient thermal injury. There have been no injuries reported as a result of this issue..

Recalling firm
GE Medical Systems, LLC
Classification
Class II
Status
Terminated
Initiated
2015-12-07
Terminated
2018-11-08
Firm location
Waukesha, WI, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

GE Healthcare, 3.0T 6 Channel Phased Array Flex Coil, M0050SS. Product Usage The 3.0T GE 6-Channel Phased Array Flex Coil is indicated for use on the order of the physician in conjunction with a 3.0T MRI scanner, as an accessory to produce 2D and 3D images.

Reason for recall

Coil overheating can occur when the device is used in Mode 2 setup. This could lead to a serious patient thermal injury. There have been no injuries reported as a result of this issue.