Recall Z-0552-2024— GE Hangwei Medical Systems, Co. LTD

Class II · Ongoing

Class II FDA medical device recall by GE Hangwei Medical Systems, Co. LTD, initiated 2023-11-07. It names one FDA 510(k) clearance: K191699. Product: The System is designed as a modular system with components that include an Overhead Tube Suspension with tube/collimator, wallstand, Table, X-ray generator, and cleared wireless digital detectors. The list of detectors verified and validated for use with the Discovery XR656 HD system, including their specifications, are provided in the user documentation. The System generates diagnostic The components may be combined in different configurations to meet specific customer needs. In addition, upgrade configurations are available for predicate devices. The optional image pasting function enables the operator to stitch sequentially acquired radiographs into a single image. This 510(k) is to incorporate the VolumeRad advanced application that was currently available on the Discovery XR656 product onto the Discovery XR656 HD, as well as introduce a new Metal Artifact Reduction Algorithm, and an optional standalone console to take any HelixTM acquired images via DICOM (such as from a Discovery XR656 HD, Optima XR646 HD, or Optima XR240amx) and process the images independently of the system it was acquired on.. Reason: GE HealthCare has become aware of the potential that incorrect torque has been applied to certain bolts of the Definium Tempo and Definium Tempo Pro products which may result in a potential fall of elements of the Overhead Tube Suspension (OTS) assembly..

Recalling firm
GE Hangwei Medical Systems, Co. LTD
Classification
Class II
Status
Ongoing
Initiated
2023-11-07
Firm location
Beijing, N/A, China

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

The System is designed as a modular system with components that include an Overhead Tube Suspension with tube/collimator, wallstand, Table, X-ray generator, and cleared wireless digital detectors. The list of detectors verified and validated for use with the Discovery XR656 HD system, including their specifications, are provided in the user documentation. The System generates diagnostic The components may be combined in different configurations to meet specific customer needs. In addition, upgrade configurations are available for predicate devices. The optional image pasting function enables the operator to stitch sequentially acquired radiographs into a single image. This 510(k) is to incorporate the VolumeRad advanced application that was currently available on the Discovery XR656 product onto the Discovery XR656 HD, as well as introduce a new Metal Artifact Reduction Algorithm, and an optional standalone console to take any HelixTM acquired images via DICOM (such as from a Discovery XR656 HD, Optima XR646 HD, or Optima XR240amx) and process the images independently of the system it was acquired on.

Reason for recall

GE HealthCare has become aware of the potential that incorrect torque has been applied to certain bolts of the Definium Tempo and Definium Tempo Pro products which may result in a potential fall of elements of the Overhead Tube Suspension (OTS) assembly.

Data sourced from openFDA. This site is unofficial and independent of the FDA.