Recall Z-0549-2018— Zimmer Biomet, Inc.
Class II · Terminated
Class II FDA medical device recall by Zimmer Biomet, Inc., initiated 2017-06-05, terminated 2019-04-17. FDA's recall record names no specific 510(k) clearance for this event. Product: Humeral Guide Wire Ball Tip 2.4 mm Diameter 70 cm Length, Item Number 47-2255-008-00 Product Usage: The Ball Tip and Tear Drop Guide Wires are used during the initial reaming of the intramedullary canal. They also assists in guiding the nail during implantation. The guide wire incorporates either a ball or tapered tip to aid in removal of the intramedullary reamer if it becomes lodged or fails.. Reason: The design verification for the previous packaging configuration G928 does not cover the 70cm wires. A design verification has been completed to move the 70cm guide wires to a new packaging configuration. As a result, the products packaged in the previous packaging configuration are being removed..
- Recalling firm
- Zimmer Biomet, Inc.
- Classification
- Class II
- Status
- Terminated
- Initiated
- 2017-06-05
- Terminated
- 2019-04-17
- Firm location
- Warsaw, IN, United States
Product description
Humeral Guide Wire Ball Tip 2.4 mm Diameter 70 cm Length, Item Number 47-2255-008-00 Product Usage: The Ball Tip and Tear Drop Guide Wires are used during the initial reaming of the intramedullary canal. They also assists in guiding the nail during implantation. The guide wire incorporates either a ball or tapered tip to aid in removal of the intramedullary reamer if it becomes lodged or fails.
Reason for recall
The design verification for the previous packaging configuration G928 does not cover the 70cm wires. A design verification has been completed to move the 70cm guide wires to a new packaging configuration. As a result, the products packaged in the previous packaging configuration are being removed.
Data sourced from openFDA. This site is unofficial and independent of the FDA.