Recall Z-0548-2018— Zimmer Biomet, Inc.

Class II · Terminated

Class II FDA medical device recall by Zimmer Biomet, Inc., initiated 2017-06-05, terminated 2019-04-17. FDA's recall record names no specific 510(k) clearance for this event. Product: M/DN Intramedullary Fixation Humeral Guide Wire - Bullet Tip 2.4 mm Diameter 70 cm Length Single Use Only, Item Number 00-2255-026-00 Product Usage: The Ball Tip and Tear Drop Guide Wires are used during the initial reaming of the intramedullary canal. They also assists in guiding the nail during implantation. The guide wire incorporates either a ball or tapered tip to aid in removal of the intramedullary reamer if it becomes lodged or fails.. Reason: The design verification for the previous packaging configuration G928 does not cover the 70cm wires. A design verification has been completed to move the 70cm guide wires to a new packaging configuration. As a result, the products packaged in the previous packaging configuration are being removed..

Recalling firm
Zimmer Biomet, Inc.
Classification
Class II
Status
Terminated
Initiated
2017-06-05
Terminated
2019-04-17
Firm location
Warsaw, IN, United States

Product description

M/DN Intramedullary Fixation Humeral Guide Wire - Bullet Tip 2.4 mm Diameter 70 cm Length Single Use Only, Item Number 00-2255-026-00 Product Usage: The Ball Tip and Tear Drop Guide Wires are used during the initial reaming of the intramedullary canal. They also assists in guiding the nail during implantation. The guide wire incorporates either a ball or tapered tip to aid in removal of the intramedullary reamer if it becomes lodged or fails.

Reason for recall

The design verification for the previous packaging configuration G928 does not cover the 70cm wires. A design verification has been completed to move the 70cm guide wires to a new packaging configuration. As a result, the products packaged in the previous packaging configuration are being removed.

Data sourced from openFDA. This site is unofficial and independent of the FDA.