Recall Z-0533-2016— Ameditech Inc

Class III · Terminated

Class III FDA medical device recall by Ameditech Inc, initiated 2015-11-16, terminated 2016-11-09. FDA's recall record names no specific 510(k) clearance for this event. Product: ProScreen 8 Panel Dip Card w/Adult, Item No. PSDA-8P, PSDA-8MBO300. Reason: Ameditechs Drugs of Abuse Tests listed above have shown reduced reactivity for up to two (2) of the nineteen (19) Benzodiazepine (BZO) compounds for which specificity information is provided in the product insert. The particular BZO compounds for which there may be reduced reactivity are Clonazepam and Chlordiazepoxide..

Recalling firm
Ameditech Inc
Classification
Class III
Status
Terminated
Initiated
2015-11-16
Terminated
2016-11-09
Firm location
San Diego, CA, United States

Product description

ProScreen 8 Panel Dip Card w/Adult, Item No. PSDA-8P, PSDA-8MBO300

Reason for recall

Ameditechs Drugs of Abuse Tests listed above have shown reduced reactivity for up to two (2) of the nineteen (19) Benzodiazepine (BZO) compounds for which specificity information is provided in the product insert. The particular BZO compounds for which there may be reduced reactivity are Clonazepam and Chlordiazepoxide.