Recall Z-0530-2020— Zimmer Biomet, Inc.
Class II · Terminated
Class II FDA medical device recall by Zimmer Biomet, Inc., initiated 2019-08-21, terminated 2021-07-21. FDA's recall record names no specific 510(k) clearance for this event. Product: BALL NOSE GUIDE WIRE 80CM, 100CM. Item Nos. 281001080 281001100 Product Usage: These instruments and delivery systems are used to facilitate implantation of orthopedic medical devices.. Reason: Elevated levels of bacterial endotoxin and residual debris remain on the devices due to cleaning issue..
- Recalling firm
- Zimmer Biomet, Inc.
- Classification
- Class II
- Status
- Terminated
- Initiated
- 2019-08-21
- Terminated
- 2021-07-21
- Firm location
- Warsaw, IN, United States
Product description
BALL NOSE GUIDE WIRE 80CM, 100CM. Item Nos. 281001080 281001100 Product Usage: These instruments and delivery systems are used to facilitate implantation of orthopedic medical devices.
Reason for recall
Elevated levels of bacterial endotoxin and residual debris remain on the devices due to cleaning issue.
Data sourced from openFDA. This site is unofficial and independent of the FDA.