Recall Z-0527-2025— Smiths Medical ASD, Inc.

Class II · Ongoing

Class II FDA medical device recall by Smiths Medical ASD, Inc., initiated 2024-10-28. FDA's recall record names no specific 510(k) clearance for this event. Product: smiths medical portex Venn Reusable Tracheal Tube Introducer, 15Ch x 60cm, REF 14-504-17 and REF 14-504-17JP. Reason: Smiths Medical has identified a potential for ingress of fluid into the device during reprocessing. The ingress takes place at the rear of the device between the end cap and the tube. This could lead to staining of the device or allowing the fluid to remain in the device. In addition, the Hypochlorite Solution (200ppm) and the 4% Acetic Acid disinfectants recommended in the IFU may be inadequate according to the disinfection standards for this type of device..

Recalling firm
Smiths Medical ASD, Inc.
Classification
Class II
Status
Ongoing
Initiated
2024-10-28
Firm location
Minneapolis, MN, United States

Product description

smiths medical portex Venn Reusable Tracheal Tube Introducer, 15Ch x 60cm, REF 14-504-17 and REF 14-504-17JP

Reason for recall

Smiths Medical has identified a potential for ingress of fluid into the device during reprocessing. The ingress takes place at the rear of the device between the end cap and the tube. This could lead to staining of the device or allowing the fluid to remain in the device. In addition, the Hypochlorite Solution (200ppm) and the 4% Acetic Acid disinfectants recommended in the IFU may be inadequate according to the disinfection standards for this type of device.