Recall Z-0527-2014— Karl Storz Endoscopy America Inc
Class II · Terminated
Class II FDA medical device recall by Karl Storz Endoscopy America Inc, initiated 2013-11-13, terminated 2014-03-27. It names 3 FDA 510(k) clearances: K981458, K954797, K832010. Product: Broncho Flexible Fiberscope, models: 11001BN1, 11002BD1, 11004BC1, and 11009BC1. " Examination and treatment of the tracheobronchial tree Contraindications for Bronchoscopy " Uncontrolled angina, uncontrolled heart failure, or serious uncontrolled ventricular arrhythmias " Uncontrolled hypertension " Bleeding disorder or use of anticoagulants.. Reason: KARL STORZ Endoscopy America, Inc. is recalling the Flexible Broncho Fiberscope or a Flexible Intubation Fiberscope with a suction port because the material used in the suction channel portion was thermoplastic polyurethane (TPU), which is not listed in STERRAD systems claims for interior endoscope lumens..
- Recalling firm
- Karl Storz Endoscopy America Inc
- Classification
- Class II
- Status
- Terminated
- Initiated
- 2013-11-13
- Terminated
- 2014-03-27
- Firm location
- Southbridge, MA, United States
510(k) clearances named in this recall
3 on FDA's record- K981458 — KSEA RHINO-LARYNGO-BRONCHO-FIBERSCOPE,PEDIATRIC BRONCHO-FIBERSCOPE, AND BRONCHO-FIBERSCOPE,11001RK, 11002BC & 110
- K954797 — KARL STORZ RHINO-LARYNGO-BRONCHO FIBERSCOPE, FIBERSCOPE FOR BRONCHOSCOPY
- K832010 — BRONCHO-FIBERSCOPE
K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
Broncho Flexible Fiberscope, models: 11001BN1, 11002BD1, 11004BC1, and 11009BC1. " Examination and treatment of the tracheobronchial tree Contraindications for Bronchoscopy " Uncontrolled angina, uncontrolled heart failure, or serious uncontrolled ventricular arrhythmias " Uncontrolled hypertension " Bleeding disorder or use of anticoagulants.
Reason for recall
KARL STORZ Endoscopy America, Inc. is recalling the Flexible Broncho Fiberscope or a Flexible Intubation Fiberscope with a suction port because the material used in the suction channel portion was thermoplastic polyurethane (TPU), which is not listed in STERRAD systems claims for interior endoscope lumens.