Recall Z-0520-2026— Medline Industries, LP
Class II · Ongoing
Class II FDA medical device recall by Medline Industries, LP, initiated 2025-09-30. FDA's recall record names no specific 510(k) clearance for this event. Product: MEDLINE convenience kits labeled as: 1) MINOR SINGLE BASIN PACK, REF DYNJ37698L; 2) BASIN PACK, REF DYNJ50123D; 3) MAJOR BASIN PACK, REF DYNJ59816A; 4) OSC BASIN PACK, REF DYNJ68217A.. Reason: Medline Industries, LP is issuing a recall for specific item(s) and lot(s) of Medline Kits containing SafeAir Smoke Evacuation Pencils. Stryker Instruments has expanded the scope of their recall of SafeAir Smoke Evacuation Pencils previously initiated in October 2024, adding an additional 47 lots to the scope of the recall. The initial recall was due to the potential for product to activate without manual input when plugged into a power source or remain active after buttons are released..
- Recalling firm
- Medline Industries, LP
- Classification
- Class II
- Status
- Ongoing
- Initiated
- 2025-09-30
- Firm location
- Northfield, IL, United States
Product description
MEDLINE convenience kits labeled as: 1) MINOR SINGLE BASIN PACK, REF DYNJ37698L; 2) BASIN PACK, REF DYNJ50123D; 3) MAJOR BASIN PACK, REF DYNJ59816A; 4) OSC BASIN PACK, REF DYNJ68217A.
Reason for recall
Medline Industries, LP is issuing a recall for specific item(s) and lot(s) of Medline Kits containing SafeAir Smoke Evacuation Pencils. Stryker Instruments has expanded the scope of their recall of SafeAir Smoke Evacuation Pencils previously initiated in October 2024, adding an additional 47 lots to the scope of the recall. The initial recall was due to the potential for product to activate without manual input when plugged into a power source or remain active after buttons are released.
Data sourced from openFDA. This site is unofficial and independent of the FDA.