Recall Z-0519-2022— GE Healthcare, LLC
Class II · Ongoing
Class II FDA medical device recall by GE Healthcare, LLC, initiated 2021-11-04. It names 2 FDA 510(k) clearances: K180995, K161588. Product: GE Vscan Extend, being sold as "Vscan Extend Dual DICOM". Reason: The firm identified that the Vscan Extend device will not power on because of depleted battery, due to the device not being turned off prior to battery removal or when being connected to a computer without AC power..
- Recalling firm
- GE Healthcare, LLC
- Classification
- Class II
- Status
- Ongoing
- Initiated
- 2021-11-04
- Firm location
- Waukesha, WI, United States
510(k) clearances named in this recall
2 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
GE Vscan Extend, being sold as "Vscan Extend Dual DICOM"
Reason for recall
The firm identified that the Vscan Extend device will not power on because of depleted battery, due to the device not being turned off prior to battery removal or when being connected to a computer without AC power.
Data sourced from openFDA. This site is unofficial and independent of the FDA.