Recall Z-0516-2018— Zimmer Biomet, Inc.
Class II · Terminated
Class II FDA medical device recall by Zimmer Biomet, Inc., initiated 2017-06-12, terminated 2018-06-29. FDA's recall record names no specific 510(k) clearance for this event. Product: DRILL BIT MINI QUICK CONNECT 1.0MM, REF 2312-20-200 & 2312-20-202, NON STERILE. Reason: A raw material anomaly was discovered in four lots of raw material and has the potential to be on or below the surface of the device components which can increase the risk of instrument fracture.
- Recalling firm
- Zimmer Biomet, Inc.
- Classification
- Class II
- Status
- Terminated
- Initiated
- 2017-06-12
- Terminated
- 2018-06-29
- Firm location
- Warsaw, IN, United States
Product description
DRILL BIT MINI QUICK CONNECT 1.0MM, REF 2312-20-200 & 2312-20-202, NON STERILE
Reason for recall
A raw material anomaly was discovered in four lots of raw material and has the potential to be on or below the surface of the device components which can increase the risk of instrument fracture