Recall Z-0515-2018— Siemens Healthcare Diagnostics, Inc

Class II · Terminated

Class II FDA medical device recall by Siemens Healthcare Diagnostics, Inc, initiated 2017-06-07, terminated 2019-05-08. It names one FDA 510(k) clearance: K100293. Product: ADVIA Centaur Calibrator 30, 2 pack, 6 pack, and Ref 2 pack. Siemens Material Number (SMN): 10379810, 10379811, 10697756.. Reason: Siemens Healthcare Diagnostics has confirmed a negative bias for ADVIA Centaur Enhanced Estradiol (eE2) on the ADVIA Centaur, ADVIA Centaur XP and ADVIA Centaur XPT systems when calibrating with Calibrator 30 kit lots ending in 21 (C3021) as compared to Calibrator 30 kit lots ending in 20 (C3020)..

Recalling firm
Siemens Healthcare Diagnostics, Inc
Classification
Class II
Status
Terminated
Initiated
2017-06-07
Terminated
2019-05-08
Firm location
East Walpole, MA, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

ADVIA Centaur Calibrator 30, 2 pack, 6 pack, and Ref 2 pack. Siemens Material Number (SMN): 10379810, 10379811, 10697756.

Reason for recall

Siemens Healthcare Diagnostics has confirmed a negative bias for ADVIA Centaur Enhanced Estradiol (eE2) on the ADVIA Centaur, ADVIA Centaur XP and ADVIA Centaur XPT systems when calibrating with Calibrator 30 kit lots ending in 21 (C3021) as compared to Calibrator 30 kit lots ending in 20 (C3020).