Recall Z-0512-2021— Talladium Inc

Class III · Terminated

Class III FDA medical device recall by Talladium Inc, initiated 2020-10-09, terminated 2022-08-17. It names one FDA 510(k) clearance: K142371. Product: Luminesse ML PMMA, 98x20mm, REF: 5336-B1 - Product Usage: used as a milling blank in the fabrication of provisional crowns and bridges. All blanks are solely used by a Trained Dental Technician or on the order of a dental professional.. Reason: Heat-cured polymethylmethacrylate (PMMA) milling blank product used to make long-term temporary crowns and bridge restorations has incorrect label. label states Part no. 5336-B1 (98x20mm). However, the item inside the box was Part no. 4302-B1 (95x20mm). An incorrect label may result in compatibility issues when setting up the disc with milling machine fixtures due to size variation..

Recalling firm
Talladium Inc
Classification
Class III
Status
Terminated
Initiated
2020-10-09
Terminated
2022-08-17
Firm location
Valencia, CA, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

Luminesse ML PMMA, 98x20mm, REF: 5336-B1 - Product Usage: used as a milling blank in the fabrication of provisional crowns and bridges. All blanks are solely used by a Trained Dental Technician or on the order of a dental professional.

Reason for recall

Heat-cured polymethylmethacrylate (PMMA) milling blank product used to make long-term temporary crowns and bridge restorations has incorrect label. label states Part no. 5336-B1 (98x20mm). However, the item inside the box was Part no. 4302-B1 (95x20mm). An incorrect label may result in compatibility issues when setting up the disc with milling machine fixtures due to size variation.

Data sourced from openFDA. This site is unofficial and independent of the FDA.