Recall Z-0512-2013— Siemens Healthcare Diagnostics, Inc
Class II · Terminated
Class II FDA medical device recall by Siemens Healthcare Diagnostics, Inc, initiated 2012-11-14, terminated 2015-09-25. It names one FDA 510(k) clearance: K982680. Product: Siemens Healthcare ADVIA Centaur Theophylline 2 (THE02) 250 Test 08838753 50 Test 03009619 Intended Use For in vitro diagnostic use in the quantitative determination of theophylline in serum or plasma using the ADVIA Centaur and ADVIA Centaur XP systems.. Reason: Operating range of the ADVIA Centaur systems is 18¿C to 30¿C (64.4¿F to 86.0¿F), a change in room temperature may increase or decrease assay results, depending on the specific assay and ADVIA Centaur system used..
- Recalling firm
- Siemens Healthcare Diagnostics, Inc
- Classification
- Class II
- Status
- Terminated
- Initiated
- 2012-11-14
- Terminated
- 2015-09-25
- Firm location
- East Walpole, MA, United States
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
Siemens Healthcare ADVIA Centaur Theophylline 2 (THE02) 250 Test 08838753 50 Test 03009619 Intended Use For in vitro diagnostic use in the quantitative determination of theophylline in serum or plasma using the ADVIA Centaur and ADVIA Centaur XP systems.
Reason for recall
Operating range of the ADVIA Centaur systems is 18¿C to 30¿C (64.4¿F to 86.0¿F), a change in room temperature may increase or decrease assay results, depending on the specific assay and ADVIA Centaur system used.
Data sourced from openFDA. This site is unofficial and independent of the FDA.