Recall Z-0510-2013— Siemens Healthcare Diagnostics, Inc
Class II · Terminated
Class II FDA medical device recall by Siemens Healthcare Diagnostics, Inc, initiated 2012-11-14, terminated 2015-09-25. It names one FDA 510(k) clearance: K970546. Product: Siemens Healthcare ADVIA Centaur Digitoxin (DGTN) 250 Test 09046419 50 Test 08861968 Intended Use For in vitro diagnostic use in the quantitative determination of digitoxin in serum or plasma using the ADVIA Centaur and ADVIA Centaur XP systems.. Reason: Operating range of the ADVIA Centaur systems is 18¿C to 30¿C (64.4¿F to 86.0¿F), a change in room temperature may increase or decrease assay results, depending on the specific assay and ADVIA Centaur system used..
- Recalling firm
- Siemens Healthcare Diagnostics, Inc
- Classification
- Class II
- Status
- Terminated
- Initiated
- 2012-11-14
- Terminated
- 2015-09-25
- Firm location
- East Walpole, MA, United States
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
Siemens Healthcare ADVIA Centaur Digitoxin (DGTN) 250 Test 09046419 50 Test 08861968 Intended Use For in vitro diagnostic use in the quantitative determination of digitoxin in serum or plasma using the ADVIA Centaur and ADVIA Centaur XP systems.
Reason for recall
Operating range of the ADVIA Centaur systems is 18¿C to 30¿C (64.4¿F to 86.0¿F), a change in room temperature may increase or decrease assay results, depending on the specific assay and ADVIA Centaur system used.