Recall Z-0509-2026— Medtronic Perfusion Systems
Class II · Ongoing
Class II FDA medical device recall by Medtronic Perfusion Systems, initiated 2025-10-03. It names 2 FDA 510(k) clearances: K250199, K230364. Product: MC3 VitalFlow Console, REF 58100; Blood pump of ecmo. Reason: As of August 6, 2025, Medtronic has received eleven reports of VitalFlow Consoles displaying an E70 error code during normal operation. When this occurs, the touch screen may become temporarily unresponsive and go blank for up to two minutes before recovering to full functionality. Importantly, the console continues to maintain set pump speed and function throughout..
- Recalling firm
- Medtronic Perfusion Systems
- Classification
- Class II
- Status
- Ongoing
- Initiated
- 2025-10-03
- Firm location
- Brooklyn Park, MN, United States
510(k) clearances named in this recall
2 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
MC3 VitalFlow Console, REF 58100; Blood pump of ecmo
Reason for recall
As of August 6, 2025, Medtronic has received eleven reports of VitalFlow Consoles displaying an E70 error code during normal operation. When this occurs, the touch screen may become temporarily unresponsive and go blank for up to two minutes before recovering to full functionality. Importantly, the console continues to maintain set pump speed and function throughout.