Recall Z-0509-2013— Siemens Healthcare Diagnostics, Inc
Class II · Terminated
Class II FDA medical device recall by Siemens Healthcare Diagnostics, Inc, initiated 2012-11-14, terminated 2015-09-25. FDA's recall record names no specific 510(k) clearance for this event. Product: Siemens Healthcare ADVIA Centaur DHEA-504 (DHEAS) 50 Test 06489701 The ADVIA Centaur DHEAS assay is an in vitro diagnostic immunoassay for the quantitative determination of dehydroepiandrosterone sulfate (DHEAS) in human serum and plasma using the ADVIA Centaur and ADVIA Centaur XP systems. Reason: Operating range of the ADVIA Centaur systems is 18C to 30¿C (64.4¿F to 86.0¿F), a change in room temperature may increase or decrease assay results, depending on the specific assay and ADVIA Centaur system used..
- Recalling firm
- Siemens Healthcare Diagnostics, Inc
- Classification
- Class II
- Status
- Terminated
- Initiated
- 2012-11-14
- Terminated
- 2015-09-25
- Firm location
- East Walpole, MA, United States
Product description
Siemens Healthcare ADVIA Centaur DHEA-504 (DHEAS) 50 Test 06489701 The ADVIA Centaur DHEAS assay is an in vitro diagnostic immunoassay for the quantitative determination of dehydroepiandrosterone sulfate (DHEAS) in human serum and plasma using the ADVIA Centaur and ADVIA Centaur XP systems
Reason for recall
Operating range of the ADVIA Centaur systems is 18C to 30¿C (64.4¿F to 86.0¿F), a change in room temperature may increase or decrease assay results, depending on the specific assay and ADVIA Centaur system used.