Recall Z-0508-2026— Thoratec LLC
Class II · Ongoing
Class II FDA medical device recall by Thoratec LLC, initiated 2025-10-09. FDA's recall record names no specific 510(k) clearance for this event. Product: Brand Name: HeartMate Product Name: HeartMate II¿ Controller (standalone) Model/Catalog Number: 106017, HeartMate II¿ Pocket Controller, US; 106762, HeartMate II¿ Pocket Controller, US; 107801, HeartMate II¿ Pocket Controller Software Version: N/A Product Description: The HeartMate II System Controller is a small computer that controls and monitors system operation. It sends power and operating signals to the Left Ventricular Assist Device and adjusts device operation to maintain programmed levels of cardiac support. The System Controller identifies alarm conditions and initiates Hazard and Advisory alarms. It records device performance and alarm data, and transfers the information to the HeartMate Touch Communication System. Component: Yes, HeartMate II¿ LVAS Implant Kit. Reason: Due to potential corrosion issues associated with the Backup Battery cable connector that may cause a Backup Battery Fault Alarm..
- Recalling firm
- Thoratec LLC
- Classification
- Class II
- Status
- Ongoing
- Initiated
- 2025-10-09
- Firm location
- Pleasanton, CA, United States
Product description
Brand Name: HeartMate Product Name: HeartMate II¿ Controller (standalone) Model/Catalog Number: 106017, HeartMate II¿ Pocket Controller, US; 106762, HeartMate II¿ Pocket Controller, US; 107801, HeartMate II¿ Pocket Controller Software Version: N/A Product Description: The HeartMate II System Controller is a small computer that controls and monitors system operation. It sends power and operating signals to the Left Ventricular Assist Device and adjusts device operation to maintain programmed levels of cardiac support. The System Controller identifies alarm conditions and initiates Hazard and Advisory alarms. It records device performance and alarm data, and transfers the information to the HeartMate Touch Communication System. Component: Yes, HeartMate II¿ LVAS Implant Kit
Reason for recall
Due to potential corrosion issues associated with the Backup Battery cable connector that may cause a Backup Battery Fault Alarm.
Data sourced from openFDA. This site is unofficial and independent of the FDA.