Recall Z-0508-2020— Medtronic Neuromodulation

Class I · Terminated

Class I FDA medical device recall by Medtronic Neuromodulation, initiated 2019-10-11, terminated 2023-09-15. FDA's recall record names no specific 510(k) clearance for this event. Product: Synchromed II programmable pump, Model numbers 8637-20 and 8637-40. Reason: There is a potential for the presence of a foreign particle inside the pump motor assembly which could interfere with the motor gear rotation and lead to a permanent motor stall..

Recalling firm
Medtronic Neuromodulation
Classification
Class I
Status
Terminated
Initiated
2019-10-11
Terminated
2023-09-15
Firm location
Minneapolis, MN, United States

Product description

Synchromed II programmable pump, Model numbers 8637-20 and 8637-40

Reason for recall

There is a potential for the presence of a foreign particle inside the pump motor assembly which could interfere with the motor gear rotation and lead to a permanent motor stall.

Data sourced from openFDA. This site is unofficial and independent of the FDA.