Recall Z-0507-2026— Thoratec LLC

Class II · Ongoing

Class II FDA medical device recall by Thoratec LLC, initiated 2025-10-09. FDA's recall record names no specific 510(k) clearance for this event. Product: Brand Name: HeartMate Product Name: HeartMate II¿ LVAS Implant Kit Model/Catalog Number: 106015, HeartMate II¿ LVAS Kit, with Pocket Controller, Sealed, US; 106016, HeartMate II¿ LVAS Kit, with Pocket Controller, Sealed, EU Software Version: N/A Product Description: The HeartMate II LVAS is an axial-flow, rotary ventricular assist system that generates flows up to 10 liters per minute (lpm). One end of the Left Ventricular Assist Device is attached to the apex of the left ventricle; the other end of the device connects to the ascending aorta. The Left Ventricular Assist Device diverts blood from the weakened left ventricle and propels it to the aorta. A small external computer, the System Controller, monitors system operation. A driveline, which passes through the patient s abdomen, connects the implanted pump to the System Controller. The system is powered by a Power Module or Mobile Power Unit (MPU) that is connected to an AC electrical outlet, or by two HeartMate 14 Volt Lithium-Ion batteries. Component: No. Reason: Due to potential corrosion issues associated with the Backup Battery cable connector that may cause a Backup Battery Fault Alarm..

Recalling firm
Thoratec LLC
Classification
Class II
Status
Ongoing
Initiated
2025-10-09
Firm location
Pleasanton, CA, United States

Product description

Brand Name: HeartMate Product Name: HeartMate II¿ LVAS Implant Kit Model/Catalog Number: 106015, HeartMate II¿ LVAS Kit, with Pocket Controller, Sealed, US; 106016, HeartMate II¿ LVAS Kit, with Pocket Controller, Sealed, EU Software Version: N/A Product Description: The HeartMate II LVAS is an axial-flow, rotary ventricular assist system that generates flows up to 10 liters per minute (lpm). One end of the Left Ventricular Assist Device is attached to the apex of the left ventricle; the other end of the device connects to the ascending aorta. The Left Ventricular Assist Device diverts blood from the weakened left ventricle and propels it to the aorta. A small external computer, the System Controller, monitors system operation. A driveline, which passes through the patient s abdomen, connects the implanted pump to the System Controller. The system is powered by a Power Module or Mobile Power Unit (MPU) that is connected to an AC electrical outlet, or by two HeartMate 14 Volt Lithium-Ion batteries. Component: No

Reason for recall

Due to potential corrosion issues associated with the Backup Battery cable connector that may cause a Backup Battery Fault Alarm.

Data sourced from openFDA. This site is unofficial and independent of the FDA.