Recall Z-0506-2025— Zyno Medical LLC
Class II · Ongoing
Class II FDA medical device recall by Zyno Medical LLC, initiated 2024-10-18. It names 2 FDA 510(k) clearances: K130690, K100705. Product: Z-800 Infusion System. Model/Catalog Number: Z-800, Z-800F, Z-800W, Z-800WF. The Z-800 Infusion system is intended to provide intravenous infusion of parenteral fluids, blood, and blood products to a patient under the direction or supervision of a physician or other certified healthcare professional.. Reason: Potential for the battery to deplete over time from self-discharge and internal parasitic leakage current despite being continuously plugged into AC mains..
- Recalling firm
- Zyno Medical LLC
- Classification
- Class II
- Status
- Ongoing
- Initiated
- 2024-10-18
- Firm location
- Natick, MA, United States
510(k) clearances named in this recall
2 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
Z-800 Infusion System. Model/Catalog Number: Z-800, Z-800F, Z-800W, Z-800WF. The Z-800 Infusion system is intended to provide intravenous infusion of parenteral fluids, blood, and blood products to a patient under the direction or supervision of a physician or other certified healthcare professional.
Reason for recall
Potential for the battery to deplete over time from self-discharge and internal parasitic leakage current despite being continuously plugged into AC mains.