Recall Z-0499-2019— Siemens Healthcare Diagnostics, Inc.

Class II · Terminated

Class II FDA medical device recall by Siemens Healthcare Diagnostics, Inc., initiated 2018-10-11, terminated 2020-06-29. FDA's recall record names no specific 510(k) clearance for this event. Product: Dimension Vista Lactate Dehydrogenase (LDI) Flex reagent cartridge, an in vitro diagnostic test for the quantitative measurement of lactate dehydrogenase in human serum and plasma on the Dimension Vista¿ System.Cat. No. K2054, Mat. No. 10464323. Reason: Siemens Healthcare Diagnostics has confirmed that Dimension Vista Lactate Dehydrogenase (LDI) may exhibit an erroneously elevated result when the preceding assay in the reaction cuvette is Lactic Acid (LA). The magnitude of the elevation may be variable. Both quality control and patient samples may be impacted. Incomplete removal of residual lactate dehydrogenase, a component of the LA reagent, has been identified as the cause of the potentially elevated result. A falsely elevated result for lactate dehydrogenase may lead to additional investigations to determine the etiology of tissue damage. Results would be correlated with clinical history and presentation in addition to other diagnostic laboratory testing..

Recalling firm
Siemens Healthcare Diagnostics, Inc.
Classification
Class II
Status
Terminated
Initiated
2018-10-11
Terminated
2020-06-29
Firm location
Newark, DE, United States

Product description

Dimension Vista Lactate Dehydrogenase (LDI) Flex reagent cartridge, an in vitro diagnostic test for the quantitative measurement of lactate dehydrogenase in human serum and plasma on the Dimension Vista¿ System.Cat. No. K2054, Mat. No. 10464323

Reason for recall

Siemens Healthcare Diagnostics has confirmed that Dimension Vista Lactate Dehydrogenase (LDI) may exhibit an erroneously elevated result when the preceding assay in the reaction cuvette is Lactic Acid (LA). The magnitude of the elevation may be variable. Both quality control and patient samples may be impacted. Incomplete removal of residual lactate dehydrogenase, a component of the LA reagent, has been identified as the cause of the potentially elevated result. A falsely elevated result for lactate dehydrogenase may lead to additional investigations to determine the etiology of tissue damage. Results would be correlated with clinical history and presentation in addition to other diagnostic laboratory testing.

Data sourced from openFDA. This site is unofficial and independent of the FDA.