Recall Z-0492-2019— Becton Dickinson & Company

Class II · Terminated

Class II FDA medical device recall by Becton Dickinson & Company, initiated 2018-08-30, terminated 2020-10-21. FDA's recall record names no specific 510(k) clearance for this event. Product: BD Vacutainer Specimen Collection Cups, Cat. No. 364975. Reason: BD Vacutainer Urine Collection Cups may contribute to a slight elevation of Red Blood Cell (RBC) counts when used with the Sysmex UF Series Analyzers. BD recommends following Good Clinical Practice (GCP) by using microscopy to confirm an elevated RBC Count in a Urine sample as per CLSI GP16-3A. 29(4) and CAP Checklist on Urinalysis. Product will continue to ship to not impact critical patient care..

Recalling firm
Becton Dickinson & Company
Classification
Class II
Status
Terminated
Initiated
2018-08-30
Terminated
2020-10-21
Firm location
Franklin Lakes, NJ, United States

Product description

BD Vacutainer Specimen Collection Cups, Cat. No. 364975

Reason for recall

BD Vacutainer Urine Collection Cups may contribute to a slight elevation of Red Blood Cell (RBC) counts when used with the Sysmex UF Series Analyzers. BD recommends following Good Clinical Practice (GCP) by using microscopy to confirm an elevated RBC Count in a Urine sample as per CLSI GP16-3A. 29(4) and CAP Checklist on Urinalysis. Product will continue to ship to not impact critical patient care.

Data sourced from openFDA. This site is unofficial and independent of the FDA.