Recall Z-0490-2021— Geistlich Pharma North America, Inc.
Class II · Terminated
Class II FDA medical device recall by Geistlich Pharma North America, Inc., initiated 2020-10-19, terminated 2021-07-13. It names one FDA 510(k) clearance: K120601. Product: Geistlich Bio-Oss Pen 0.25-1mm 0.25g - Product Usage: natural bone mineral for filling of bone defects in maxillofacial surgery, implantology, and periodontology. Product Number: 20115. Reason: Difficult to expel Geistlich Bio-Oss granules from the pen, resistance could suddenly drop, and the granules may be expelled all at once, causing surrounding tissue, particularly in use for the sinus lift indication and while unlikely, could pose a risk for patient harm, including irritation of the oral mucosa or even a perforation of the Schneiderian membrane.
- Recalling firm
- Geistlich Pharma North America, Inc.
- Classification
- Class II
- Status
- Terminated
- Initiated
- 2020-10-19
- Terminated
- 2021-07-13
- Firm location
- Princeton, NJ, United States
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
Geistlich Bio-Oss Pen 0.25-1mm 0.25g - Product Usage: natural bone mineral for filling of bone defects in maxillofacial surgery, implantology, and periodontology. Product Number: 20115
Reason for recall
Difficult to expel Geistlich Bio-Oss granules from the pen, resistance could suddenly drop, and the granules may be expelled all at once, causing surrounding tissue, particularly in use for the sinus lift indication and while unlikely, could pose a risk for patient harm, including irritation of the oral mucosa or even a perforation of the Schneiderian membrane
Data sourced from openFDA. This site is unofficial and independent of the FDA.