Recall Z-0489-2014— GE Healthcare, LLC

Class II · Terminated

Class II FDA medical device recall by GE Healthcare, LLC, initiated 2013-10-11, terminated 2016-02-03. It names 3 FDA 510(k) clearances: K130584, K122253, K120598. Product: HGE Healthcare Patient Monitor B30, Procare Monitor B20, Procare Monitor B40, and Monitor B40 K122253 (Procare Monitor B20): The PROCARE Monitor B20 is a portable multiparameter unit to be used for monitoring and recording of, and to generate alarms for, multiple physiological parameters of adult, pediatric, and neonatal patients in a hospital environment and during intra-hospital transport. The PROCARE Monitor B20 is intended for use under the direct supervision of a licensed health care practitioner. The PROCARE Monitor B20 is not intended for use during MRI. The PROCARE Monitor B20 monitors and displays oscillometric non-invasive blood pressure (systolic, diastolic and mean arterial pressure), invasive blood pressure, end-tidal carbon dioxide, heart/pulse rate, respiration rate, EGG (including arrhythmia and ST segment analysis), temperature with a reusable or disposable electronic thermometer for continual monitoring. Esosophageal/Nasopharyngeal/Tymnpanic/Rectal/Bladder/Axillary/Skin/Airway/Room/Myo cardial/Core/Surface temperature, and functional oxygen saturation (Sp0 2) and pulse rate via continuous monitoring , including monitoring during conditions of clinical patient motion or low perfusion. K120598 (Procare Monitor B40): The PROCARE Monitor 840 is a portable multiparameter unit to be used for monitoring and recording of, and to generate alarms for, multiple physiological parameters of adult, pediatric, and neonatal patients in a hospital environment and during intra-hospital transport. The PROCARE Monitor 840 is intended for use under the direct supervision of a licensed health care practitioner. The PROCARE Monitor 840 is not intended for use during MRI. The PROCARE Monitor B40 monitors and displays oscillometric non-invasive blood pressure (systolic, diastolic and mean arterial pressure), invasive blood pressure, end-tidal carbon dioxide, heart/pulse rate, respiration rate, ECG (including arrhythmia and ST segment analysis), temperature with a reusable or disposable electronic thermometer for continual monitoring. Esophageal/Nasopharyngeal/Tympaiic/Rectal/Bladder/Axillary/Skin/Airway/Room/Myocardial/Core/ Surface temperature, and functional oxygen saturation (SpO2) and pulse rate via continuous monitoring, including monitoring during conditions of clinical patient motion or low perfusion. K130584 (Monitor B40): The Monitor B40 is a portable multi-parameter unit to be used for monitoring and recording of, and to generate alarms for, multiple physiological parameters of adult, pediatric, and neonatal patients in a hospital environment and during intra-hospital transport. The Monitor B40 is intended for use under the direct supervision of a licensed health care practitioner. The Monitor B40 is not intended for use during MRI. The Monitor B40 can be a stand-alone monitor or interfaced to other devices via a network. The Monitor B40 monitors and displays: ECG (including ST segment, arrhythmia detection), invasive blood pressure, heart/pulse rate, oscillometric non-invasive blood pressure (systolic, diastolic and mean arterial pressure), functional oxygen saturation (SpO2) and pulse rate via continuous monitoring (including monitoring during conditions of clinical patient motion or low perfusion), temperature with a reusable or disposable electronic thermometer for continual monitoring. Esophageal/Nasopharyngeal/Tympanic/Rectal/Bladder/Axillary/Skin/Airway/Room/Myocardial/Core/ Surface temperature, impedance respiration, respiration rate, airway gases (CO2, O2, N2O, anesthetic agents, anesthetic agent identification and respiratory rate) and Entropy. Product Quantity Distributed (Int) 16,061 (343 US, 15,718 OUS). Reason: GE Healthcare has recently become aware of a potential safety issue due to ECG filter settings associated with the B20/B40 V1, B30 and B40 V2 Patient Monitors. The B20, B30 and B40 Patient Monitors set the ECG filter setting incorrectly if the monitor starts with impedance respiration set in the ON position. Instead of the user selected filter, an additional 0.5 Hz high pass filter is used. The.

Recalling firm
GE Healthcare, LLC
Classification
Class II
Status
Terminated
Initiated
2013-10-11
Terminated
2016-02-03
Firm location
Waukesha, WI, United States

510(k) clearances named in this recall

3 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

HGE Healthcare Patient Monitor B30, Procare Monitor B20, Procare Monitor B40, and Monitor B40 K122253 (Procare Monitor B20): The PROCARE Monitor B20 is a portable multiparameter unit to be used for monitoring and recording of, and to generate alarms for, multiple physiological parameters of adult, pediatric, and neonatal patients in a hospital environment and during intra-hospital transport. The PROCARE Monitor B20 is intended for use under the direct supervision of a licensed health care practitioner. The PROCARE Monitor B20 is not intended for use during MRI. The PROCARE Monitor B20 monitors and displays oscillometric non-invasive blood pressure (systolic, diastolic and mean arterial pressure), invasive blood pressure, end-tidal carbon dioxide, heart/pulse rate, respiration rate, EGG (including arrhythmia and ST segment analysis), temperature with a reusable or disposable electronic thermometer for continual monitoring. Esosophageal/Nasopharyngeal/Tymnpanic/Rectal/Bladder/Axillary/Skin/Airway/Room/Myo cardial/Core/Surface temperature, and functional oxygen saturation (Sp0 2) and pulse rate via continuous monitoring , including monitoring during conditions of clinical patient motion or low perfusion. K120598 (Procare Monitor B40): The PROCARE Monitor 840 is a portable multiparameter unit to be used for monitoring and recording of, and to generate alarms for, multiple physiological parameters of adult, pediatric, and neonatal patients in a hospital environment and during intra-hospital transport. The PROCARE Monitor 840 is intended for use under the direct supervision of a licensed health care practitioner. The PROCARE Monitor 840 is not intended for use during MRI. The PROCARE Monitor B40 monitors and displays oscillometric non-invasive blood pressure (systolic, diastolic and mean arterial pressure), invasive blood pressure, end-tidal carbon dioxide, heart/pulse rate, respiration rate, ECG (including arrhythmia and ST segment analysis), temperature with a reusable or disposable electronic thermometer for continual monitoring. Esophageal/Nasopharyngeal/Tympaiic/Rectal/Bladder/Axillary/Skin/Airway/Room/Myocardial/Core/ Surface temperature, and functional oxygen saturation (SpO2) and pulse rate via continuous monitoring, including monitoring during conditions of clinical patient motion or low perfusion. K130584 (Monitor B40): The Monitor B40 is a portable multi-parameter unit to be used for monitoring and recording of, and to generate alarms for, multiple physiological parameters of adult, pediatric, and neonatal patients in a hospital environment and during intra-hospital transport. The Monitor B40 is intended for use under the direct supervision of a licensed health care practitioner. The Monitor B40 is not intended for use during MRI. The Monitor B40 can be a stand-alone monitor or interfaced to other devices via a network. The Monitor B40 monitors and displays: ECG (including ST segment, arrhythmia detection), invasive blood pressure, heart/pulse rate, oscillometric non-invasive blood pressure (systolic, diastolic and mean arterial pressure), functional oxygen saturation (SpO2) and pulse rate via continuous monitoring (including monitoring during conditions of clinical patient motion or low perfusion), temperature with a reusable or disposable electronic thermometer for continual monitoring. Esophageal/Nasopharyngeal/Tympanic/Rectal/Bladder/Axillary/Skin/Airway/Room/Myocardial/Core/ Surface temperature, impedance respiration, respiration rate, airway gases (CO2, O2, N2O, anesthetic agents, anesthetic agent identification and respiratory rate) and Entropy. Product Quantity Distributed (Int) 16,061 (343 US, 15,718 OUS)

Reason for recall

GE Healthcare has recently become aware of a potential safety issue due to ECG filter settings associated with the B20/B40 V1, B30 and B40 V2 Patient Monitors. The B20, B30 and B40 Patient Monitors set the ECG filter setting incorrectly if the monitor starts with impedance respiration set in the ON position. Instead of the user selected filter, an additional 0.5 Hz high pass filter is used. The