Recall Z-0485-2019— Tosoh Bioscience Inc

Class II · Terminated

Class II FDA medical device recall by Tosoh Bioscience Inc, initiated 2018-09-28, terminated 2020-07-08. FDA's recall record names no specific 510(k) clearance for this event. Product: Tosoh AIA-900 immunoassay Analyzer. Reason: A software issue may cause the liquid volume detection entries for 2nd and 3rd reagent bottles to be overwritten with the data from the 1st reagent bottle. Tosoh has released software version 2.51 (or later) to fix the problem. This failure is mitigated by the hardware design. The consequence would be a slight delay in testing patient sample or reporting patient result..

Recalling firm
Tosoh Bioscience Inc
Classification
Class II
Status
Terminated
Initiated
2018-09-28
Terminated
2020-07-08
Firm location
Grove City, OH, United States

Product description

Tosoh AIA-900 immunoassay Analyzer

Reason for recall

A software issue may cause the liquid volume detection entries for 2nd and 3rd reagent bottles to be overwritten with the data from the 1st reagent bottle. Tosoh has released software version 2.51 (or later) to fix the problem. This failure is mitigated by the hardware design. The consequence would be a slight delay in testing patient sample or reporting patient result.