Recall Z-0485-2014— Abbott Diabetes Care, Inc.

Class I · Terminated

Class I FDA medical device recall by Abbott Diabetes Care, Inc., initiated 2013-11-18, terminated 2015-08-03. It names one FDA 510(k) clearance: K092602. Product: FreeStyle Blood Glucose Test Strips; For in vitro diagnostic testing. 50 count UPC 6 99073 12050 2; Freestyle Blood Glucose Test Strips; for Institutional Use only 50 count: UPC 6 99073 70792 5 Manufactured by Abbott Diabetes Care Inc. Alameda, CA. The FreeStyle Blood Glucose test strip is intended for use in the quantitative measurement of glucose in capillary whole blood from the finger, upper arm and palm. It is intended for use by healthcare professionals and people with diabetes mellitus at home as an aid in monitoring the effectiveness of a diabetes control program. It is not intended for the diagnosis of or screening for diabetes mellitus, and it is not intended for use on neonates or arterial blood.. Reason: Certain lots of FreeStyle and FreeStyle Lite Blood Glucose Test Strips produce erroneously low blood glucose results when using FreeStyle Blood Glucose Meters, FreeStyle Flash Blood Glucose Meters and the FreeStyle blood glucose meter built into the OmniPod system. Erroneously low results that are not recognized may pose significant risks to your health..

Recalling firm
Abbott Diabetes Care, Inc.
Classification
Class I
Status
Terminated
Initiated
2013-11-18
Terminated
2015-08-03
Firm location
Alameda, CA, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

FreeStyle Blood Glucose Test Strips; For in vitro diagnostic testing. 50 count UPC 6 99073 12050 2; Freestyle Blood Glucose Test Strips; for Institutional Use only 50 count: UPC 6 99073 70792 5 Manufactured by Abbott Diabetes Care Inc. Alameda, CA. The FreeStyle Blood Glucose test strip is intended for use in the quantitative measurement of glucose in capillary whole blood from the finger, upper arm and palm. It is intended for use by healthcare professionals and people with diabetes mellitus at home as an aid in monitoring the effectiveness of a diabetes control program. It is not intended for the diagnosis of or screening for diabetes mellitus, and it is not intended for use on neonates or arterial blood.

Reason for recall

Certain lots of FreeStyle and FreeStyle Lite Blood Glucose Test Strips produce erroneously low blood glucose results when using FreeStyle Blood Glucose Meters, FreeStyle Flash Blood Glucose Meters and the FreeStyle blood glucose meter built into the OmniPod system. Erroneously low results that are not recognized may pose significant risks to your health.