Recall Z-0478-2021— RAYSEARCH LABORATORIES AB

Class II · Terminated

Class II FDA medical device recall by RAYSEARCH LABORATORIES AB, initiated 2020-10-08, terminated 2023-03-30. It names one FDA 510(k) clearance: K200569. Product: RayStation standalone software treatment planning system, Model No. 4.0, 4.5, 4.7, 4.9, 5.0, 6.0, 6.1, 6.2, 6.3, 7.0, 8.0, 8.1, 9.0, 9.1, 9.2, 10.0 - Product Usage: RayStation is a software system for radiation therapy and medical oncology. Based on user input, RayStation proposes treatment plans. After a proposed treatment plan is reviewed and approved by authorized intended users, RayStation may also be used to administer treatments. The system functionality can be configured based on user needs.. Reason: If a region of interest (ROI) or point of interest (POI) that is referenced from an imported plan is missing in the imported RT Structure Set, the reference may become linked to the wrong ROI or POI..

Recalling firm
RAYSEARCH LABORATORIES AB
Classification
Class II
Status
Terminated
Initiated
2020-10-08
Terminated
2023-03-30
Firm location
Stockholm, N/A, Sweden

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

RayStation standalone software treatment planning system, Model No. 4.0, 4.5, 4.7, 4.9, 5.0, 6.0, 6.1, 6.2, 6.3, 7.0, 8.0, 8.1, 9.0, 9.1, 9.2, 10.0 - Product Usage: RayStation is a software system for radiation therapy and medical oncology. Based on user input, RayStation proposes treatment plans. After a proposed treatment plan is reviewed and approved by authorized intended users, RayStation may also be used to administer treatments. The system functionality can be configured based on user needs.

Reason for recall

If a region of interest (ROI) or point of interest (POI) that is referenced from an imported plan is missing in the imported RT Structure Set, the reference may become linked to the wrong ROI or POI.