Recall Z-0477-2013— Arjo, Inc. dba ArjoHuntleigh

Class II · Terminated

Class II FDA medical device recall by Arjo, Inc. dba ArjoHuntleigh, initiated 2012-11-16, terminated 2016-09-13. FDA's recall record names no specific 510(k) clearance for this event. Product: Enterprise 9000 Bed; an AC-powered adjustable hospital bed with built-in electronic weigh scale, patient bed exit alarm, and under bed anti-entrapment system; ArjoHuntleigh AB, Eslov, Sweden; Models 9000, 9600, 9700 and 9800 General medicine, general and specialist surgery, care of the elderly, HDU, ITU, coronary care and oncology. The bed is appropriate for high dependency patients who pose a movement and handling risk and/or whose clinical condition requires that they are positioned with the minimal physical handling.. Reason: ArjoHuntleigh has received reports of unintended movement where the backrest and/or thigh sections of the bed have moved without anyone consciously pressing the control button. These incidents have occurred in High Dependency and Intensive Therapy environments where unintended movement of the patient may have serious consequences..

Recalling firm
Arjo, Inc. dba ArjoHuntleigh
Classification
Class II
Status
Terminated
Initiated
2012-11-16
Terminated
2016-09-13
Firm location
Addison, IL, United States

Product description

Enterprise 9000 Bed; an AC-powered adjustable hospital bed with built-in electronic weigh scale, patient bed exit alarm, and under bed anti-entrapment system; ArjoHuntleigh AB, Eslov, Sweden; Models 9000, 9600, 9700 and 9800 General medicine, general and specialist surgery, care of the elderly, HDU, ITU, coronary care and oncology. The bed is appropriate for high dependency patients who pose a movement and handling risk and/or whose clinical condition requires that they are positioned with the minimal physical handling.

Reason for recall

ArjoHuntleigh has received reports of unintended movement where the backrest and/or thigh sections of the bed have moved without anyone consciously pressing the control button. These incidents have occurred in High Dependency and Intensive Therapy environments where unintended movement of the patient may have serious consequences.

Data sourced from openFDA. This site is unofficial and independent of the FDA.