Recall Z-0476-2021— Luminex Corporation

Class II · Terminated

Class II FDA medical device recall by Luminex Corporation, initiated 2020-10-13, terminated 2024-01-22. It names one FDA 510(k) clearance: K143653. Product: Verigene Processor SP, Model Numbers 10-0000-07 and 10-0000-07R. Used with assay: Luminex, VERIGENE Gram-Positive Blood Culture Nucleic Acid Test,BC-GP, IVD - Product Usage: IVD sample processor, which utilizes a single-use disposable test consumable and a self-contained Verigene Test Cartridge for each sample tested.. Reason: Luminex has determined that three (3) separate test fixtures used by field service engineers to service the VERIGENE SP Processors were procured from unapproved sources and, in some cases, were manufactured using unvalidated processes. These text fixtures were used to calibrate and verify VERIGENESP devices..

Recalling firm
Luminex Corporation
Classification
Class II
Status
Terminated
Initiated
2020-10-13
Terminated
2024-01-22
Firm location
Northbrook, IL, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

Verigene Processor SP, Model Numbers 10-0000-07 and 10-0000-07R. Used with assay: Luminex, VERIGENE Gram-Positive Blood Culture Nucleic Acid Test,BC-GP, IVD - Product Usage: IVD sample processor, which utilizes a single-use disposable test consumable and a self-contained Verigene Test Cartridge for each sample tested.

Reason for recall

Luminex has determined that three (3) separate test fixtures used by field service engineers to service the VERIGENE SP Processors were procured from unapproved sources and, in some cases, were manufactured using unvalidated processes. These text fixtures were used to calibrate and verify VERIGENESP devices.