Recall Z-0473-2021— Siemens Healthcare Diagnostics, Inc.

Class II · Terminated

Class II FDA medical device recall by Siemens Healthcare Diagnostics, Inc., initiated 2020-10-26, terminated 2024-04-19. It names one FDA 510(k) clearance: K151792. Product: Atellica IM 1300 Analyzer - Product Usage: automated, immunoassay analyzers designed to perform in vitro diagnostic tests on clinical specimens. Siemens Material Number (SMN): 11066001. Reason: Test Definition scanning may reset custom settings to defaults causing falsely low or high results. When a 2D Master Curve and TDef barcode for a new kit lot of reagent is scanned and the TDef version is a newer version than the version that is currently on the system, some of the customer defined settings for that assay may reset to default values.

Recalling firm
Siemens Healthcare Diagnostics, Inc.
Classification
Class II
Status
Terminated
Initiated
2020-10-26
Terminated
2024-04-19
Firm location
Tarrytown, NY, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

Atellica IM 1300 Analyzer - Product Usage: automated, immunoassay analyzers designed to perform in vitro diagnostic tests on clinical specimens. Siemens Material Number (SMN): 11066001

Reason for recall

Test Definition scanning may reset custom settings to defaults causing falsely low or high results. When a 2D Master Curve and TDef barcode for a new kit lot of reagent is scanned and the TDef version is a newer version than the version that is currently on the system, some of the customer defined settings for that assay may reset to default values