Recall Z-0470-2019— Synaptive Medical Inc
Class I · Terminated
Class I FDA medical device recall by Synaptive Medical Inc, initiated 2018-10-09, terminated 2021-08-12. It names one FDA 510(k) clearance: K153281. Product: BrightMatter Guide, Product Number SYN-0026, Versions: 1.5.0, 1.5.1 and 1.5.2 Product Usage: BrightMatter Guide can navigate a third-party port (NICO BrainPath device) by attaching Synaptive s custom Tracking Array to the port directly. As the BrainPath is navigable in two configurations (sheath only / sheath and obturator), the tip of the obturator is represented in the software with a tip extension beyond the bottom of the sheath. Note sheath and cannula are used interchangeably in this document.. Reason: This recall has been initiated due to a software defect found in the Guide System software when used with NICO BrainPath ports under certain circumstances. When one trajectory is set with a blue port and another trajectory is set with a gold port, the software defect is triggered when the user switches between these trajectories during a surgical procedure..
- Recalling firm
- Synaptive Medical Inc
- Classification
- Class I
- Status
- Terminated
- Initiated
- 2018-10-09
- Terminated
- 2021-08-12
- Firm location
- Toronto, N/A, Canada
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
BrightMatter Guide, Product Number SYN-0026, Versions: 1.5.0, 1.5.1 and 1.5.2 Product Usage: BrightMatter Guide can navigate a third-party port (NICO BrainPath device) by attaching Synaptive s custom Tracking Array to the port directly. As the BrainPath is navigable in two configurations (sheath only / sheath and obturator), the tip of the obturator is represented in the software with a tip extension beyond the bottom of the sheath. Note sheath and cannula are used interchangeably in this document.
Reason for recall
This recall has been initiated due to a software defect found in the Guide System software when used with NICO BrainPath ports under certain circumstances. When one trajectory is set with a blue port and another trajectory is set with a gold port, the software defect is triggered when the user switches between these trajectories during a surgical procedure.