Recall Z-0468-2015— Summit Industries Inc.
Class II · Terminated
Class II FDA medical device recall by Summit Industries Inc., initiated 2014-10-28, terminated 2015-05-21. It names one FDA 510(k) clearance: K850198. Product: The J700 Floor Mounted Tube Stand (FMTS) is intended to support and position the diagnostic x-ray tube housing assembly for a medical radiographic procedure. The product consists of a vertical column (with counterbalanced vertical carriage, counterweight and pulley system), horizontal floor car and a floor mounted track to ride on.. Reason: The welds on J700 tube stands may be insufficient resulting in tube stand breakage, damage to the X-Ray machine, and/or patient or healthcare provider injury..
- Recalling firm
- Summit Industries Inc.
- Classification
- Class II
- Status
- Terminated
- Initiated
- 2014-10-28
- Terminated
- 2015-05-21
- Firm location
- Chicago, IL, United States
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
The J700 Floor Mounted Tube Stand (FMTS) is intended to support and position the diagnostic x-ray tube housing assembly for a medical radiographic procedure. The product consists of a vertical column (with counterbalanced vertical carriage, counterweight and pulley system), horizontal floor car and a floor mounted track to ride on.
Reason for recall
The welds on J700 tube stands may be insufficient resulting in tube stand breakage, damage to the X-Ray machine, and/or patient or healthcare provider injury.