Recall Z-0462-2025— Getinge Usa Sales Inc
Class II · Ongoing
Class II FDA medical device recall by Getinge Usa Sales Inc, initiated 2024-10-03. FDA's recall record names no specific 510(k) clearance for this event. Product: MEERA CL mobile operating table - Designed for the placement and positioning of the patient immediately before, during and after surgical interventions, as well as for examination and treatment Item Number: 700001B0. Reason: Error code "50037" occurs on the IR-Hand Control when a MEERA-table is controlled, and the table stops the movement. This error may occur sporadically when the MEERA-table is controlled via an IR-Hand Control and may result in procedural delay.
- Recalling firm
- Getinge Usa Sales Inc
- Classification
- Class II
- Status
- Ongoing
- Initiated
- 2024-10-03
- Firm location
- Wayne, NJ, United States
Product description
MEERA CL mobile operating table - Designed for the placement and positioning of the patient immediately before, during and after surgical interventions, as well as for examination and treatment Item Number: 700001B0
Reason for recall
Error code "50037" occurs on the IR-Hand Control when a MEERA-table is controlled, and the table stops the movement. This error may occur sporadically when the MEERA-table is controlled via an IR-Hand Control and may result in procedural delay
Data sourced from openFDA. This site is unofficial and independent of the FDA.