Recall Z-0459-2016— Siemens Healthcare Diagnostics Inc.
Class II · Terminated
Class II FDA medical device recall by Siemens Healthcare Diagnostics Inc., initiated 2015-11-10, terminated 2017-08-23. It names one FDA 510(k) clearance: K130276. Product: Reagent Management System (RMS) Dimension EXL with LM STM Product Usage: The Reagent Management System (RMS) is a module for use with the Dimension¿ clinical chemistry system. The RMS doubles the available storage of refrigerated Flex¿ reagent cartridges on the Dimension¿ system from 44 to 88. The RMS loads, hydrates, and automatically removes Flex¿ reagent cartridges, allowing the Dimension¿ system to process samples without interruption.. Reason: Siemens has identified RMS refrigeration compressors that are missing the safety cover on the electrical termination block. The exposed electrical hazard on the electrical termination block is located directly behind the RMS Waste Container..
- Recalling firm
- Siemens Healthcare Diagnostics Inc.
- Classification
- Class II
- Status
- Terminated
- Initiated
- 2015-11-10
- Terminated
- 2017-08-23
- Firm location
- Brookfield, CT, United States
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
Reagent Management System (RMS) Dimension EXL with LM STM Product Usage: The Reagent Management System (RMS) is a module for use with the Dimension¿ clinical chemistry system. The RMS doubles the available storage of refrigerated Flex¿ reagent cartridges on the Dimension¿ system from 44 to 88. The RMS loads, hydrates, and automatically removes Flex¿ reagent cartridges, allowing the Dimension¿ system to process samples without interruption.
Reason for recall
Siemens has identified RMS refrigeration compressors that are missing the safety cover on the electrical termination block. The exposed electrical hazard on the electrical termination block is located directly behind the RMS Waste Container.
Data sourced from openFDA. This site is unofficial and independent of the FDA.