Recall Z-0455-2016— CareFusion 303, Inc.
Class II · Terminated
Class II FDA medical device recall by CareFusion 303, Inc., initiated 2015-11-12, terminated 2017-01-23. It names one FDA 510(k) clearance: K031741. Product: Alaris EtCO2 Module, Model 8300. A capnograph that continuously monitors end tidal carbon dioxide (EtCO2), fractional inspired carbon dioxide (FiCO2), and respiratory rate (RR).. Reason: CareFusion is recalling the Alaris EtCO2 module model 8300 because it may have been tested during service using a higher than specified concentration of CO2. The effect to the EtCO2 module tested with a higher concentration of CO2 could lead to an incorrect measurement of CO2..
- Recalling firm
- CareFusion 303, Inc.
- Classification
- Class II
- Status
- Terminated
- Initiated
- 2015-11-12
- Terminated
- 2017-01-23
- Firm location
- San Diego, CA, United States
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
Alaris EtCO2 Module, Model 8300. A capnograph that continuously monitors end tidal carbon dioxide (EtCO2), fractional inspired carbon dioxide (FiCO2), and respiratory rate (RR).
Reason for recall
CareFusion is recalling the Alaris EtCO2 module model 8300 because it may have been tested during service using a higher than specified concentration of CO2. The effect to the EtCO2 module tested with a higher concentration of CO2 could lead to an incorrect measurement of CO2.