Recall Z-0455-2015— ConMed Corporation

Class I · Terminated

Class I FDA medical device recall by ConMed Corporation, initiated 2014-11-06, terminated 2017-10-30. It names one FDA 510(k) clearance: K014209. Product: ADULT Radiotransparent Electrode, REF/Catalog Number 2001H-C, Rx ONLY --- COMPARABLE TO PHILIPS MEDICAL HEARTSTART ELECTRODES PLUG STYLE CONNECTOR. Reason: Philips Healthcare made changes to the design of the connection between multifunction electrodes and their HeartStart FR3 and HeartStart FRx AEDs. The failure to connect may result in a delay of therapy, which could result in patient death or serious injury..

Recalling firm
ConMed Corporation
Classification
Class I
Status
Terminated
Initiated
2014-11-06
Terminated
2017-10-30
Firm location
Utica, NY, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

ADULT Radiotransparent Electrode, REF/Catalog Number 2001H-C, Rx ONLY --- COMPARABLE TO PHILIPS MEDICAL HEARTSTART ELECTRODES PLUG STYLE CONNECTOR

Reason for recall

Philips Healthcare made changes to the design of the connection between multifunction electrodes and their HeartStart FR3 and HeartStart FRx AEDs. The failure to connect may result in a delay of therapy, which could result in patient death or serious injury.

Data sourced from openFDA. This site is unofficial and independent of the FDA.