Recall Z-0454-2026— GE Healthcare Finland Oy
Class II · Ongoing
Class II FDA medical device recall by GE Healthcare Finland Oy, initiated 2025-09-26. It names one FDA 510(k) clearance: K213234. Product: CARESCAPE ONE, Model no. 2087075-300 - shipped with potentially affected batteries or affected batteries installed as a spare part Monitor, Physiological, Patient (With Arrhythmia Detection or Alarms Updated 3/19 to add CARESCAPE ONE MBZ101 2087075-001. Reason: Potential loss of monitoring on affected monitors when powered by certain batteries. Affected batteries can lose capability to hold charge and provide back up power. If this situation occurs, it will cause loss of patient monitoring and potentially delay recognizing patient status changes requiring treatment..
- Recalling firm
- GE Healthcare Finland Oy
- Classification
- Class II
- Status
- Ongoing
- Initiated
- 2025-09-26
- Firm location
- Helsinki, Finland
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
CARESCAPE ONE, Model no. 2087075-300 - shipped with potentially affected batteries or affected batteries installed as a spare part Monitor, Physiological, Patient (With Arrhythmia Detection or Alarms Updated 3/19 to add CARESCAPE ONE MBZ101 2087075-001
Reason for recall
Potential loss of monitoring on affected monitors when powered by certain batteries. Affected batteries can lose capability to hold charge and provide back up power. If this situation occurs, it will cause loss of patient monitoring and potentially delay recognizing patient status changes requiring treatment.
Data sourced from openFDA. This site is unofficial and independent of the FDA.