Recall Z-0454-2016— DePuy Spine, Inc.

Class II · Terminated

Class II FDA medical device recall by DePuy Spine, Inc., initiated 2015-11-20, terminated 2017-07-03. FDA's recall record names no specific 510(k) clearance for this event. Product: Aegis Torque Handle CATALOG NO. : 2871-10-800. A spinal surgical instrument used on the AEGIS Anterior Lumbar Plate System.. Reason: Manufacturing error within the Torque Handle may prohibit some handles from being able to fully assemble with the tightener shaft. The design of the inspection gage does not account for the additional shift length.

Recalling firm
DePuy Spine, Inc.
Classification
Class II
Status
Terminated
Initiated
2015-11-20
Terminated
2017-07-03
Firm location
Raynham, MA, United States

Product description

Aegis Torque Handle CATALOG NO. : 2871-10-800. A spinal surgical instrument used on the AEGIS Anterior Lumbar Plate System.

Reason for recall

Manufacturing error within the Torque Handle may prohibit some handles from being able to fully assemble with the tightener shaft. The design of the inspection gage does not account for the additional shift length