Recall Z-0452-2015— GE Healthcare, LLC
Class II · Terminated
Class II FDA medical device recall by GE Healthcare, LLC, initiated 2014-09-04, terminated 2015-01-21. It names one FDA 510(k) clearance: K133278. Product: Discovery IGS 740 Indicated for use for patients from newborn to geriatric in generating fluoroscopic and rotational images of human anatomy for cardiovascular, vascular and non-vascular, diagnostic and interventional procedures.. Reason: Loss of imaging - affected devices may experience a loss of real-time interventional imaging and/or gantry motion due to the Electromagnetic (EM) sensitivity of the grid detection mechanism..
- Recalling firm
- GE Healthcare, LLC
- Classification
- Class II
- Status
- Terminated
- Initiated
- 2014-09-04
- Terminated
- 2015-01-21
- Firm location
- Waukesha, WI, United States
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
Discovery IGS 740 Indicated for use for patients from newborn to geriatric in generating fluoroscopic and rotational images of human anatomy for cardiovascular, vascular and non-vascular, diagnostic and interventional procedures.
Reason for recall
Loss of imaging - affected devices may experience a loss of real-time interventional imaging and/or gantry motion due to the Electromagnetic (EM) sensitivity of the grid detection mechanism.