Recall Z-0447-2026— Medline Industries, LP

Class I · Ongoing

Class I FDA medical device recall by Medline Industries, LP, initiated 2025-10-10. FDA's recall record names no specific 510(k) clearance for this event. Product: Convenience Kit, BREAST HERNIA DYNJ909125G. Reason: Firm received multiple complaints of anesthesia circuit leaks due to cracks in the 120" expandable tubing. Cracks and resulting leaks may lead to hypoxia, hypoventilation, and inadequate delivery of inhaled anesthetic agents, which may result in serious injury or death. Volatile anesthetics may also escape into the operating room environment..

Recalling firm
Medline Industries, LP
Classification
Class I
Status
Ongoing
Initiated
2025-10-10
Firm location
Northfield, IL, United States

Product description

Convenience Kit, BREAST HERNIA DYNJ909125G

Reason for recall

Firm received multiple complaints of anesthesia circuit leaks due to cracks in the 120" expandable tubing. Cracks and resulting leaks may lead to hypoxia, hypoventilation, and inadequate delivery of inhaled anesthetic agents, which may result in serious injury or death. Volatile anesthetics may also escape into the operating room environment.