Recall Z-0447-2022— Invacare Corporation

Class II · Ongoing

Class II FDA medical device recall by Invacare Corporation, initiated 2021-11-12. It names one FDA 510(k) clearance: K202379. Product: Invacare AVIVA Storm RX with LiNX Gyro Power Wheelchair (with LiNX Gyro Module Model Numbers DLX-GRY100-A and DLX-GRY101-A) Models: SRX-20R, SRX-20SP, SRX-20MP. Reason: Power Wheelchairs with a LiNX Gyro component running firmware version 6.1.2 can experience a more aggressive deceleration rate than the programmed rate due to the system following the incorrect deceleration profile, resulting in potential injury.

Recalling firm
Invacare Corporation
Classification
Class II
Status
Ongoing
Initiated
2021-11-12
Firm location
Elyria, OH, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

Invacare AVIVA Storm RX with LiNX Gyro Power Wheelchair (with LiNX Gyro Module Model Numbers DLX-GRY100-A and DLX-GRY101-A) Models: SRX-20R, SRX-20SP, SRX-20MP

Reason for recall

Power Wheelchairs with a LiNX Gyro component running firmware version 6.1.2 can experience a more aggressive deceleration rate than the programmed rate due to the system following the incorrect deceleration profile, resulting in potential injury

Data sourced from openFDA. This site is unofficial and independent of the FDA.