Recall Z-0445-2026— Medline Industries, LP
Class I · Ongoing
Class I FDA medical device recall by Medline Industries, LP, initiated 2025-10-10. FDA's recall record names no specific 510(k) clearance for this event. Product: Anesthesia Circuit Kits and other various kits that contain Anesthesia Circuits; ENT DYNJ909128I; LARYNGOSCOPY - BRONCHOSCOPY DYNJ909114D. Reason: Firm received multiple complaints of anesthesia circuit leaks due to cracks in the 120" expandable tubing. Cracks and resulting leaks may lead to hypoxia, hypoventilation, and inadequate delivery of inhaled anesthetic agents, which may result in serious injury or death. Volatile anesthetics may also escape into the operating room environment..
- Recalling firm
- Medline Industries, LP
- Classification
- Class I
- Status
- Ongoing
- Initiated
- 2025-10-10
- Firm location
- Northfield, IL, United States
Product description
Anesthesia Circuit Kits and other various kits that contain Anesthesia Circuits; ENT DYNJ909128I; LARYNGOSCOPY - BRONCHOSCOPY DYNJ909114D
Reason for recall
Firm received multiple complaints of anesthesia circuit leaks due to cracks in the 120" expandable tubing. Cracks and resulting leaks may lead to hypoxia, hypoventilation, and inadequate delivery of inhaled anesthetic agents, which may result in serious injury or death. Volatile anesthetics may also escape into the operating room environment.
Data sourced from openFDA. This site is unofficial and independent of the FDA.