Recall Z-0445-2024— Elekta, Inc.

Class II · Ongoing

Class II FDA medical device recall by Elekta, Inc., initiated 2023-10-25. It names one FDA 510(k) clearance: K210500. Product: ELEKTA LIMITED Versa HD, REF XRT 2121, 2131; Medical Charged-Particle Radiation Therapy System Accelerator, Linear. Reason: Elekta has identified that if the microswitch within the middle arm is incorrectly setup, there is a potential for an uncontrolled extension of iViewGT / XVI detector arms. When the detector arm is fully deployed there is no risk of uncontrolled extension..

Recalling firm
Elekta, Inc.
Classification
Class II
Status
Ongoing
Initiated
2023-10-25
Firm location
Atlanta, GA, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

ELEKTA LIMITED Versa HD, REF XRT 2121, 2131; Medical Charged-Particle Radiation Therapy System Accelerator, Linear

Reason for recall

Elekta has identified that if the microswitch within the middle arm is incorrectly setup, there is a potential for an uncontrolled extension of iViewGT / XVI detector arms. When the detector arm is fully deployed there is no risk of uncontrolled extension.

Data sourced from openFDA. This site is unofficial and independent of the FDA.