Recall Z-0444-2024— Elekta, Inc.
Class II · Ongoing
Class II FDA medical device recall by Elekta, Inc., initiated 2023-10-25. It names one FDA 510(k) clearance: K210500. Product: ELEKTA LIMITED Elekta Synergy Platform, REF MRT 8501, MRT 8511; Medical Charged-Particle Radiation Therapy System Accelerator, Linear. Reason: Elekta has identified that if the microswitch within the middle arm is incorrectly setup, there is a potential for an uncontrolled extension of iViewGT / XVI detector arms. When the detector arm is fully deployed there is no risk of uncontrolled extension..
- Recalling firm
- Elekta, Inc.
- Classification
- Class II
- Status
- Ongoing
- Initiated
- 2023-10-25
- Firm location
- Atlanta, GA, United States
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
ELEKTA LIMITED Elekta Synergy Platform, REF MRT 8501, MRT 8511; Medical Charged-Particle Radiation Therapy System Accelerator, Linear
Reason for recall
Elekta has identified that if the microswitch within the middle arm is incorrectly setup, there is a potential for an uncontrolled extension of iViewGT / XVI detector arms. When the detector arm is fully deployed there is no risk of uncontrolled extension.
Data sourced from openFDA. This site is unofficial and independent of the FDA.