Recall Z-0444-2022— AMO Manufacturing USA, LLC
Class II · Ongoing
Class II FDA medical device recall by AMO Manufacturing USA, LLC, initiated 2021-10-21. It names one FDA 510(k) clearance: K113479. Product: Catalys Precision Laser System, Models: Catalys-U (United States), Catalys-I (International), Catalys-C (China); and Catalys Precision Laser System No Bed, Models: 0160-6020 (United States), 0160-6010 (International). Reason: Aberrometer and precision laser systems are affected by a remote code execution vulnerability existing when the Microsoft Windows Print Spooler service is enabled, and improperly performs privileged file operations. This could be exploited to run arbitrary code with SYSTEM privileges, allowing for program installation; view/change/deleting data; or creation of new accounts with full user rights..
- Recalling firm
- AMO Manufacturing USA, LLC
- Classification
- Class II
- Status
- Ongoing
- Initiated
- 2021-10-21
- Firm location
- Milpitas, CA, United States
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
Catalys Precision Laser System, Models: Catalys-U (United States), Catalys-I (International), Catalys-C (China); and Catalys Precision Laser System No Bed, Models: 0160-6020 (United States), 0160-6010 (International)
Reason for recall
Aberrometer and precision laser systems are affected by a remote code execution vulnerability existing when the Microsoft Windows Print Spooler service is enabled, and improperly performs privileged file operations. This could be exploited to run arbitrary code with SYSTEM privileges, allowing for program installation; view/change/deleting data; or creation of new accounts with full user rights.
Data sourced from openFDA. This site is unofficial and independent of the FDA.