Recall Z-0443-2021— Maquet Cardiovascular Us Sales, Llc
Class II · Terminated
Class II FDA medical device recall by Maquet Cardiovascular Us Sales, Llc, initiated 2020-09-30, terminated 2023-05-25. It names one FDA 510(k) clearance: K090147. Product: Blood Monitoring Unit (BMU 40), model no. 701040852 - Product Usage: is to monitor blood parameters during cardiopulmonary bypass (CPB) or similar procedures with extracorporeal circulation, which require continuous monitoring of the arterial and/or venous blood parameters: partial pressure of oxygen (pO2), temperature (TArt and TVen), oxygen saturation (SO2), hemoglobin (Hb) and hematocrit (Hct).. Reason: The Printed Circuit Board Assembly (PCBA) are not in compliance with the IEC standard (60601 1, third edition). The BMU 40 units affected by this recall were designed with a short creepage distance of 1.8 mm between the two distinct and proximal areas on the printed circuit board (PCB) where as the standard requires a creepage distance of 4 mm..
- Recalling firm
- Maquet Cardiovascular Us Sales, Llc
- Classification
- Class II
- Status
- Terminated
- Initiated
- 2020-09-30
- Terminated
- 2023-05-25
- Firm location
- Wayne, NJ, United States
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
Blood Monitoring Unit (BMU 40), model no. 701040852 - Product Usage: is to monitor blood parameters during cardiopulmonary bypass (CPB) or similar procedures with extracorporeal circulation, which require continuous monitoring of the arterial and/or venous blood parameters: partial pressure of oxygen (pO2), temperature (TArt and TVen), oxygen saturation (SO2), hemoglobin (Hb) and hematocrit (Hct).
Reason for recall
The Printed Circuit Board Assembly (PCBA) are not in compliance with the IEC standard (60601 1, third edition). The BMU 40 units affected by this recall were designed with a short creepage distance of 1.8 mm between the two distinct and proximal areas on the printed circuit board (PCB) where as the standard requires a creepage distance of 4 mm.